Published March 2005 | Version v1
Book

A true doserate μ-R survey meter for low level dose rate measurements

  • 1. Nucleonix Systems Pvt Ltd, Hyderabad (India)

Description

The need for a flat energy response and sensitive low level survey meter led to the development of μR survey meter based on tissue response plastic scintillator. This survey meter comprises of a tissue response plastic scintillator and PMT, a compact HV module, charge integrator, ADC and other digital circuitry for read out and display of dose-rate. This paper describes the design concept and reports the dose-rate response measurements carried out with this meter. (author)

Part of:
Proceedings of DAE-BRNS national symposium on compact nuclear instruments and radiation detector-2005

Additional details

Publishing Information

Publisher
Bhabha Atomic Research Centre
Imprint Place
Mumbai (India)
ISBN
81-88513-15-6
Imprint Title
Proceedings of DAE-BRNS national symposium on compact nuclear instruments and radiation detector-2005
Imprint Pagination
730 p.
Journal Page Range
p. 436-438

Conference

Title
compact nuclear instruments and radiation detectors-2005
Acronym
CNIRD-2005
Dates
2-4 Mar 2005
Place
Jodhpur (India)

Optional Information

Notes
3 figs.