Published April 1990 | Version v1
Journal article

Constrained least squares filtering in high resolution PET and SPECT imaging

  • 1. Michigan Univ., Ann Arbor, MI (USA). Div. of Nuclear Medicine

Description

Constrained least squares filtering is a technique which employs a priori tomograph response information for the filtering of projection data prior to the filtered backprojection of radionuclide distribution images. This paper describes a simulation study which evaluates the performance of this algorithm as a function of the sampling density within each projection of the radon transform. The results of this evaluation demonstrate that the efficient application of this algorithm requires higher sampling densities than are typically employed in high resolution PET and SPECT. Therefore, application of this algorithm requires software and/or hardware modification of the data acquisition sampling schemes employed in tomographic systems

Additional details

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
37
Journal Issue
2
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
647-651
ISSN
0018-9499
CODEN
IETNA

Conference

Title
Institute for Electronic and Electrical Engineers (IEEE) nuclear science symposium.
Dates
15-19 Jan 1990.
Place
San Francisco, CA (USA).

Optional Information

Secondary number(s)
CONF-900143--.