Evaluation of the impact of dental artefacts on intensity-modulated radiotherapy planning for the head and neck
- 1. North Western Medical Physics, Christie Hospital NHS Foundation Trust, Manchester (United Kingdom)
Description
Background and purpose: High density materials create severe artefacts in the computed tomography (CT) scans used for radiotherapy dose calculations. Increased use of intensity-modulated radiotherapy (IMRT) to treat oropharyngeal cancers raises concerns over the accuracy of the resulting dose calculation. This work quantifies their impact and evaluates a simple corrective technique. Materials and methods: Fifteen oropharyngeal patients with severe artefacts were retrospectively planned with IMRT using two different CT/density look-up tables. Each plan was recalculated using a corrected CT dataset to evaluate the dose distribution delivered to the patient. Plan quality in the absence of dental artefacts was similarly assessed. A range of dosimetric and radiobiological parameters were compared pre- and post-correction. Results: Plans using a standard CT/density look-up table (density ≤1.8 g/cm3) revealed inconsistent inter-patient errors, mostly within clinical acceptance, although potentially significantly reducing target coverage for individual patients. Using an extended CT/density look-up table (density ≤10.0 g/cm3) greatly reduced the errors for 13/15 patients. In 2/15 patients with residual errors the CTV extended into the severely affected region and could be corrected by applying a simple manual correction. Conclusions: Use of an extended CT/density look-up table together with a simple manual bulk density correction reduces the impact of dental artefacts on head and neck IMRT planning to acceptable levels.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.radonc.2009.10.006Additional details
Identifiers
- DOI
- 10.1016/j.radonc.2009.10.006;
- PII
- S0167-8140(09)00572-6;
Publishing Information
- Journal Title
- Radiotherapy and Oncology
- Journal Volume
- 93
- Journal Issue
- 3
- Journal Page Range
- p. 553-558
- ISSN
- 0167-8140
- CODEN
- RAONDT
INIS
- Country of Publication
- Ireland
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 41074349
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; CAT SCANNING; EVALUATION; HEAD; NECK; NEOPLASMS; PATIENTS; PLANNING; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY
- Descriptors DEC
- BODY; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DISEASES; DOSES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY; TOMOGRAPHY
Optional Information
- Copyright
- Copyright (c) 2009 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.