Intercomparison and calibration of reference radiation fields for individual dose monitoring

  • 1. CNNC Nuclear Power Operations Management Co., Ltd., Haiyan (China)
  • 2. China Institute for Radiation Protection, Taiyuan (China)

Description

An introduction is made to the methodologies in which both intercomparison and calibration are conducted for individual dose monitoring center and TLD reference radiation fields, together with the 2001-2012 inter-comparison results between QNNP and CIRP. The results indicate a bias within ± 5% between measured results and conventional true values. The problems existing in the intercomparison are analyzed and solved on the basis of the work in fist half of 2008. (authors)