Published April 1998 | Version v1
Journal article

Study of enhanced MR angiography by simulation method

  • 1. Nippon Medical School, Inba (Japan). Chiba Hokusou Hospital

Description

The influences on blood vessel images of different imaging delays after medium infusion and different k-space scanning was investigated in simulation and phantom experiments. For 3D SPGR (sequential phase order), it was found that too long a delay in the timing of blood vessel imaging resulted in a loss of data in the low frequency range of the k-space and only the edge of the blood vessel image was detected. Therefore, it seemed difficult to obtain good contrast under such conditions. From the present results, it was concluded that the most favorable blood vessel images would be obtained under MRA conditions that provide high homogeneity and good contrast during scanning in k-space. (author)

Additional details

Publishing Information

Journal Title
Nippon Hoshasen Gijutsu Gakkai Zasshi
Journal Volume
54
Journal Issue
4
Journal Page Range
p. 505-511
ISSN
0369-4305