Published July 2017
| Version v1
Journal article
The Monte Carlo method for the detective efficiency of radioxenon
- 1. Northwest Institute of Nuclear Technology, Xi'an (China)
- 2. Key Laboratory of Particle and Radiation Imaging of Ministry of Education, Department of Engineering Physics, Tsinghua University, Beijing (China)
Description
For the purpose of rapid on-site measurement for radioxenon, the efficiency of radioxenon for HPGe detector were performed with Monte Carlo method for the adsorption column in the integrative sample box which was designed with functions of adsorption and source container. And the voxel filling method was utilized to describe the complicated geometric structure of that column. The results of experimental validation indicated that the relative errors for the efficiencies between experiment and calculation are less than ± 8%, and the method presented in this paper was feasible. (authors)
Additional details
Publishing Information
- Journal Title
- Nuclear Electronics and Detection Technology
- Journal Volume
- 37
- Journal Issue
- 7
- Journal Page Range
- p. 678-682
- ISSN
- 0258-0934
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 53032784
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ADSORPTION; EFFICIENCY; ERRORS; GEOMETRY; HIGH-PURITY GE DETECTORS; MEASURING METHODS; MONTE CARLO METHOD; VALIDATION; XENON ISOTOPES
- Descriptors DEC
- CALCULATION METHODS; GE SEMICONDUCTOR DETECTORS; ISOTOPES; MATHEMATICS; MEASURING INSTRUMENTS; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; SORPTION; TESTING
Optional Information
- Notes
- 5 figs., 2 tabs., 4 refs.