Creation and clinical application of real-time dose monitor using dose area product meter
Creators
- 1. Kanazawa Univ. (Japan). Hospital
- 2. Kanazawa Univ. (Japan). School of Medicine
- 3. Shiga Medical Center for Adults, Moriyama (Japan)
Description
The management of patient dose has become more of an issue in recent years. Dose can be determined non-invasively and in real time through the use of a dose area product meter, but it is the area dose value that is obtained. Therefore, we created a program that estimates entrance skin dose (ESD) in real time from area dose values obtained during procedures. We used Microsoft Visual C++ 6.0 (Standard Edition) for the programming language and C language for the programming environment. The value was a maximum 285.4 mGy at ileus tube insertion when measuring ESD for radiography of the digestive organ and non-vascular type interventional radiology (IVR) using the created program and seeking the average according to the procedures. The program that we created can be considered valid for monitoring ESD correctly and in real time. (author)
Additional details
Publishing Information
- Journal Title
- Nippon Hoshasen Gijutsu Gakkai Zasshi
- Journal Volume
- 60
- Journal Issue
- 5
- Journal Page Range
- p. 719-724
- ISSN
- 0369-4305
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 35077760
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; CONTROL SYSTEMS; DOSEMETERS; IONIZATION CHAMBERS; PATIENTS; PERSONNEL MONITORING; PROGRAMMING; RADIATION DOSES; RADIATION MONITORS; SKIN ABSORPTION
- Descriptors DEC
- ABSORPTION; DIAGNOSTIC TECHNIQUES; DOSES; MEASURING INSTRUMENTS; MEDICINE; MONITORING; MONITORS; NUCLEAR MEDICINE; RADIATION DETECTORS; RADIATION MONITORING; RADIOLOGY; SORPTION; UPTAKE