Evaluation of left ventricular function and volume with multidetector-row computed tomography. Comparison with electrocardiogram-gated single photon emission computed tomography
- 1. Toho Univ., Omori Medical Center, Tokyo (Japan)
Description
This study compared left ventricular systolic function and volume determined by multidetector-row computed tomography (MDCT) and electrocardiogram-gated single photon emission computed tomography (G-SPECT) Thirty-seven patients with coronary artery disease and non-cardiovascular disease underwent MDCT. In this study, left ventricular ejection fraction (EF), left ventricular end-diastolic volume (EDV) and left ventricular end-systolic volume (ESV) were calculated using only two-phase imaging with MDCT. Left ventricular function and volume were compared using measurements from G-SPECT. We conducted MDCT and G-SPECT virtually simultaneously. Both the EF and ESV evaluated by MDCT closely correlated with G-SPECT (r=0.763, P<0.001 and r=0.761, P<0.001, respectively). Heart rate (>65 bpm) during MDCT significantly influenced the difference in EF calculated from MDCT and G-SPECT (P<0.05). Left ventricular function can be measured with MDCT as well as G-SPECT. However, a heart rate over 65 bpm during MDCT negatively affects the EF correlation between MDCT and G-SPECT. (author)
Additional details
Publishing Information
- Journal Title
- Nippon Gazo Igaku Zasshi
- Journal Volume
- 26
- Journal Issue
- 1
- Journal Page Range
- p. 12-20
- ISSN
- 0289-0925
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 39020919
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CAT SCANNING; COMPARATIVE EVALUATIONS; ELECTROCARDIOGRAMS; IOPAMIDOL; ISONITRILES; MYOCARDIAL INFARCTION; MYOCARDIUM; PATIENTS; SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY; TECHNETIUM 99; VOLUME
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BODY; CARBONIC ACID DERIVATIVES; CARDIOVASCULAR DISEASES; CARDIOVASCULAR SYSTEM; COMPUTERIZED TOMOGRAPHY; CONTRAST MEDIA; DIAGNOSTIC TECHNIQUES; DIAGRAMS; DISEASES; EMISSION COMPUTED TOMOGRAPHY; EVALUATION; HEART; HOURS LIVING RADIOISOTOPES; INFORMATION; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MUSCLES; NUCLEI; ODD-EVEN NUCLEI; ORGANIC COMPOUNDS; ORGANS; RADIOISOTOPES; TECHNETIUM ISOTOPES; TOMOGRAPHY; YEARS LIVING RADIOISOTOPES