Published July 2007 | Version v1
Journal article

Determination of dose calculation parameters of model 81-02 198Au brachytherapy source with Monte-Carlo method

  • 1. Tsinghua Univ., Beijing (China). Dept. of Engineering Physics
  • 2. Suzhou Univ., Suzhou (China). Dept. of Radiation Medicine and Public Health

Description

Dosimetry parameters, including dose rate constant, radial dose function and anisotropy function, of model 81-02 198Au brachytherapy source were calculated in a 30 cm radius theoretical sphere by Monte-Carlo method for clinical application. For the single source, the deviation of dose rate constant is 1.113 cGy · h-1 · U-1, and within 0.18% and 1.62% from the calculated and measured value (TLDs) of Dauffy et al. Calculated value of radial dose function, g(r), is in the rang of 0.1-10.0 cm at transverse axis. Anisotropy function, F(r, θ), is calculated in 0 degree-90 degree of θ with increment of 10 degree and 0.5-9.0 cm of r with increment of 1 cm. Empiric equations are acquired for radial dose function and anisotropy function by curve fitting. (authors)

Additional details

Publishing Information

Journal Title
Atomic Energy Science and Technology
Journal Volume
41
Journal Issue
4
Journal Page Range
p. 501-504
ISSN
1000-6931

Optional Information

Notes
2 figs., 3 tabs., 6 refs.