Published January 1994 | Version v1
Journal article

Optimization of a polarized source for in vivo x-ray fluorescence analysis of platinum and other heavy metals

Creators

  • 1. Univ. of Wales, Swansea (United Kingdom). Swansea In-Vivo Analysis Research Group

Description

The Monte Carlo method was used to optimize a polarized photon source for the x-ray fluorescence analysis of platinum and other heavy metals in vivo. The source consisted of a 140 kVp, 25 mA x-ray tube with the photons plane-polarized by 90o scattering. The use of plane-polarized photons results in a significant reduction in background when the fluorescent radiation is measured along the direction of polarization. A Monte Carlo computer programme was written to simulate the production and interaction of polarized photons in order to determine the optimal polarizing material and dimensions, together with beam width and geometrical arrangement of source, polarizer and beam collimators. Calculated photon energy distributions are compared with experimental data to test the validity of the model. (author)

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
39
Journal Issue
1
Journal Page Range
p. 197-206.
ISSN
0031-9155
CODEN
PHMBA7