Published September 1993 | Version v1
Journal article

Quantitative analysis of occluded gases in mixed oxide fuel pellets by high temperature and vacuum extraction-gas chromatography

  • 1. Power Reactor and Nuclear Fuel Development Corp., Tokai, Ibaraki (Japan). Tokai Works

Description

We have developed a technique for the quantitative analysis of an occluded gases in the sintered plutonium-uranium mixed oxide (MOX) fuel pellets. The total volume of occluded gases in MOX pellet was determined using a high temperature and vacuum extraction system. The extracted gases were transferred into a special gas sampler by the McLeod gauge and their gas components were analyzed using the gas chromatography technique. Hydrogen, oxygen, nitrogen, methane, carbondioxide and carbonmonoxide were detected in the extracted gases. Relative standard deviation (n=10) were 1.0, 3.4, 0.6, 1.0, 1.4, 0.8 and 0.9 % for hydrogen, oxygen, nitrogen, methane, carbondioxide, carbonmonoxide and ethane, respectively. The time required for one determination is less than one hour. (author)

Additional details

Publishing Information

Journal Title
Donen Giho
Journal Issue
no.87
Journal Page Range
p. 45-52.
ISSN
0289-6605
CODEN
DOGIEL