Published 1977
| Version v1
Journal article
Amperometric determination of vanadium(3) and vanadium(4) in glasses
Description
A possibility is shown of determining vanadium(3) and vanadium (4) by amperometric titration with complexon 3 when they are simultaneously present in glasses. Titration has been conducted against oxidation current of complexon 3 at a potential of +1.0 V. Phosphorus, barium, silicon, and alkaline metals do not interfere with the determination. Aluminium effect has been ruled out by introduction of sodium fluoride. The error of determination does not exceed 3%. The method of determining vanadium in glasses has been proposed on the basis of the research conducted. A number of glasses and an artificial mixture have been analyzed using the method proposed
Additional details
Additional titles
- Original title (Russian)
- Амперометрическое определение ванадия (3) и (4) в стеклах
Publishing Information
- Journal Title
- Zavod. Lab.
- Journal Volume
- 43
- Journal Issue
- 2
- Series
- Zavod. Lab.
- Journal Page Range
- 153-154
INIS
- Country of Publication
- USSR
- Country of Input or Organization
- USSR
- INIS RN
- 9356944
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Descriptors DEI
- AMPEROMETRY; GLASS; IMPURITIES; QUANTITATIVE CHEMICAL ANALYSIS; VANADIUM COMPOUNDS
- Descriptors DEC
- CHEMICAL ANALYSIS; TITRATION; TRANSITION ELEMENT COMPOUNDS
Optional Information
- Notes
- For English translation see the journal Ind. Lab.