Published October 2009 | Version v1
Journal article

Calibration of a flatbed scanner for traceable paper area measurement

  • 1. JMK Instruments Oy, Kotkankallionkatu 11, FI-70600 Kuopio (Finland)
  • 2. Centre for Metrology and Accreditation (MIKES), Tekniikantie 1, FI-02150 Espoo (Finland)

Description

In this note, we describe a method to calibrate a flatbed scanner to perform traceable paper sample area measurements. Also a detailed uncertainty analysis is presented. The achieved relative uncertainty at the 95% confidence level of the area measurement is under 0.1%. Described calibration procedures can also be used in other applications where dimensional measurements are performed with a flatbed scanner. (technical design note)

Availability note (English)

Available from http://dx.doi.org/10.1088/0957-0233/20/10/107003

Additional details

Identifiers

DOI
10.1088/0957-0233/20/10/107003;
PII
S0957-0233(09)23095-5;

Publishing Information

Journal Title
Measurement Science and Technology
Journal Volume
20
Journal Issue
10
Journal Page Range
[4 p.]
ISSN
0957-0233
CODEN
MSTCEP

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
45005399
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CALIBRATION; DESIGN; PAPER