Published January 21, 2002 | Version v1
Journal article

An algorithm for the optimum combination of data from arbitrary magnetic resonance phased array probes

  • 1. CRC Clinical MR Research Group, Institute of Cancer Research, Royal Marsden NHS Trust, Sutton, Surrey (United Kingdom)
  • 2. CRC Clinical MR Research Group, Institute of Cancer Research, Royal Marsden NHS Trust, Sutton, Surrey (GB)

Description

When summing the spectra acquired with phased array coils, signals with low signal-to-noise ratio or wrongly corrected phase may degrade the overall signal-to-noise ratio (SNR). Here we present a mathematical expression predicting the dependence of combined SNR on the signal-to-noise ratios and errors in phase correction of composite signals. Based on this equation, signals that do not lead to an overall increase in signal-to-noise ratio can be identified and excluded from the weighted sum of signals. This tool is particularly useful for the combination of large numbers of signals. Additionally, a simple and robust algorithm for calculating the complex weighting factors necessary for the signal-to-noise weighted combination of spectroscopic data is presented. Errors in the calculation and correction of relative phase differences between composite spectra are analysed. The errors have a negligible effect on the overall spectral SNR for typical clinical magnetic resonance spectroscopy (MRS). The signal combination routine developed here has been applied to the first in vivo MRS study of human rectal adenocarcinomas at 1.5 T (Dzik-Jurasz A S K, Murphy P S, George M, Prock T, Collins D J, Swift I and Leach M O 2001 Magn. Reson. Med. at press), showing improvements of combined spectral SNR of up to 34% over the maximum SNR from a single element. (author)

Availability note (English)

Available online at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
47
Journal Issue
2
Journal Page Range
p. N39-N46
ISSN
0031-9155

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
33021048
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ALGORITHMS; DIAGNOSTIC USES; NMR IMAGING; NUCLEAR MAGNETIC RESONANCE; SIGNAL-TO-NOISE RATIO
Descriptors DEC
DIAGNOSTIC TECHNIQUES; MAGNETIC RESONANCE; MATHEMATICAL LOGIC; RESONANCE; USES