Measuring PET scanner sensitivity

  • 1. Neurology Dept., Memorial Sloan-Kettering Cancer Center, New York, NY (USA)
  • 2. CTI PET Systems Inc., Knoxville, TN (USA)
  • 3. California Univ., Los Angeles, CA (USA). Nuclear Medicine Lab.

Description

Sensitivity parameters derived from a plot of a scanner's true coincidence count (TCC) rates as a function of activity in a 20 cm cylindrical phantom have no direct link to image quality. Noise equivalent count (NEC) rate curves, which incorporate the noise effects of subtracting the randoms and scatter count components provide a direct link between image signal-to-noise ratios and the scatter, randoms and trues coincidence count rates. The authors have measured TCC and NEC curves with a standardized 20 cm diameter nylon cylinder for five different PET scanners with several scanner-collimator combinations. In addition, the authors have compared TCC and NEC curves on one scanner with those from an Alderson brain phantom