Published September 1991 | Version v1
Journal article

Response analysis for an approximate 3-D image reconstruction in cone-beam SPECT

  • 1. National Inst. of Radiological Sciences, Chiba (Japan)

Description

Cone-beam single photon emission computed tomography (SPECT) offers the potential for a large increase in sensitivity as compared with parallel hole or fan-beam collimation. Three-dimensional image reconstruction was approximately accomplished by backprojecting filtered projections using a two-dimensional fan-beam algorithm. The cone-beam projection data were formed from mathematical phantoms as analytically derived line integrals of the density. In order to reduce the processing time, the filtered projections were backprojected into each plane parallel to the circle on which the focal point moved. Discrepancy of source position and degradation of resolution were investigated by computer simulation in three-dimensional image space. The results obtained suggest that, the nearer to the central plane or the axis of rotation, the less image degradation is performed. By introducing a parameter of angular difference between the focal point and a fixed point in the image space during rotation, degradation of the reconstructed image can be estimated for any cone-beam SPECT system. (author)

Additional details

Publishing Information

Journal Title
Kaku Igaku
Journal Volume
28
Journal Issue
9
Series
Kaku Igaku.
Journal Page Range
1041-1047
ISSN
0022-7854
CODEN
KAIGB

INIS