Response analysis for an approximate 3-D image reconstruction in cone-beam SPECT
Creators
- 1. National Inst. of Radiological Sciences, Chiba (Japan)
Description
Cone-beam single photon emission computed tomography (SPECT) offers the potential for a large increase in sensitivity as compared with parallel hole or fan-beam collimation. Three-dimensional image reconstruction was approximately accomplished by backprojecting filtered projections using a two-dimensional fan-beam algorithm. The cone-beam projection data were formed from mathematical phantoms as analytically derived line integrals of the density. In order to reduce the processing time, the filtered projections were backprojected into each plane parallel to the circle on which the focal point moved. Discrepancy of source position and degradation of resolution were investigated by computer simulation in three-dimensional image space. The results obtained suggest that, the nearer to the central plane or the axis of rotation, the less image degradation is performed. By introducing a parameter of angular difference between the focal point and a fixed point in the image space during rotation, degradation of the reconstructed image can be estimated for any cone-beam SPECT system. (author)
Additional details
Publishing Information
- Journal Title
- Kaku Igaku
- Journal Volume
- 28
- Journal Issue
- 9
- Series
- Kaku Igaku.
- Journal Page Range
- 1041-1047
- ISSN
- 0022-7854
- CODEN
- KAIGB
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 23022370
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALGORITHMS; COLLIMATORS; CONES; IMAGE PROCESSING; IMAGES; PHANTOMS; ROTATION; SINGLE PHOTON ECT; THREE-DIMENSIONAL CALCULATIONS
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; EMISSION COMPUTED TOMOGRAPHY; MOCKUP; STRUCTURAL MODELS; TOMOGRAPHY