Delay controlled multichannel unit for registration of signals from a scintillation hodoscope

Description

Delay controlled multichannel unit for registration of signals going via 200 m coaxial cable from the ALPHA setup scintillation hogoscope is descibed. Along with the functions of signal amplification and forming by 18 channels of data transfer the unit provides for the manual control via key switch the signal delay for each channel within 0 to 15 ns by 3 steps. As delay elements microcircuits of 138 series are used. The input and output signal levels are ECL. The unit was tested and introduced into the ALPHA setup apparatus complex during physical investigations on the deuteron beam of JINR synchrophasotron