Eight years of IMRT quality assurance with ionization chambers and film dosimetry: experience of the montpellier comprehensive cancer center
Creators
- 1. Département de Cancérologie Radiothérapie et de Radiophysique, CRLC Val d'Aurelle-Paul Lamarque, Montpellier (France)
- 2. Département de Radio-Oncologie, Hôpital Maisonneuve-Rosemont, Montréal (Canada)
Description
To present the results of quality assurance (QA) in IMRT of film dosimetry and ionization chambers measurements with an eight year follow-up. All treatment plans were validated under the linear accelerator by absolute and relative measures obtained with ionization chambers (IC) and with XomatV and EDR2 films (Kodak). The average difference between IC measured and computed dose at isocenter with the gantry angle of 0° was 0.07 ± 1.22% (average ± 1 SD) for 2316 prostate, 1.33 ± 3.22% for 808 head and neck (h&n), and 0.37 ± 0.62% for 108 measurements of prostate bed fields. Pelvic treatment showed differences of 0.49 ± 1.86% in 26 fields for prostate cases and 2.07 ± 2.83% in 109 fields of anal canal. Composite measurement at isocenter for each patient showed an average difference with computed dose of 0.05 ± 0.87% for 386 prostate, 1.49 ± 1.86% for 158 h&n, 0.37 ± 0.34% for 23 prostate bed, 0.80 ± 0.28% for 4 pelvis, and 2.31 ± 0.56% for 17 anal canal cases. On the first 250 h&n analyzed by film in absolute dose, the average of the points crossing a gamma index 3% and 3 mm was 93%. This value reached 99% for the prostate fields. More than 3500 beams were found to be within the limits defined as validated for treatment between 2001 and 2008
Availability note (English)
Available from http://dx.doi.org/10.1186/1748-717X-6-85; Available from http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3155475Additional details
Identifiers
Publishing Information
- Journal Title
- Radiation Oncology (Online)
- Journal Volume
- 6
- Journal Page Range
- p. 85
- ISSN
- 1748-717X
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 47061721
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- FILM DOSIMETRY; LINEAR ACCELERATORS; PROSTATE; QUALITY ASSURANCE; RADIATION DOSES; RADIOTHERAPY
- Descriptors DEC
- ACCELERATORS; BODY; DOSES; DOSIMETRY; GLANDS; MALE GENITALS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; THERAPY
Optional Information
- Copyright
- Copyright (c)2011 Fenoglietto et al
- Notes