Published October 1987 | Version v1
Journal article

The advantage of anti-Compton counting in the measurement of low-level radioactivity by gamma-ray spetrometry

Creators

  • 1. Netherlands Energy Research Foundation, Petten

Description

The inherent advantage of anti-Compton counting of low-level radioactivity is defined in terms of changes in the limits of decision, detection and determination. It appears that the advantage factor can be expressed in the experimentally observed reductions of peak area, Compton continuum and natural background. For an array equipped with a sample changer, the advantage factor for interference by the Compton continuum is ≤4, while that for interference by the natural background amounts to ∼3. By sacrificing the sample changer and closing its gap with a NaI-plug, additional factors of ∼1.5 and ∼2 can be obtained for Compton continuum and natural background, respectively. (author) 8 refs., 4 figs

Additional details

Publishing Information

Journal Title
J. Radioanal. Nucl. Chem.
Journal Volume
115
Journal Issue
1
Series
J. Radioanal. Nucl. Chem. ;Articles.
Journal Page Range
159-173
ISSN
0236-5731
CODEN
JRNCD

Conference

Title
1. International conference on low-level measurements of actinides and long-lived radionuclides in biological and environmental samples.
Dates
9-13 Jun 1986.
Place
Lund (Sweden).