Optimization of collimators for gamma cameras with timing modulation of a signal

Creators

  • 1. Moskovskij Inzhenerno-Fizicheskij Inst. (USSR)

Description

Collimators optimized according to the static image quality criterion using in gamma cameras with timing modulation of a signal is described. Optimization has been carried out on the (0,1) matrix class describing channels arrangement in the collimator. The advantage of integral introscopes in comparison with the traditional ones is shown. The results obtained can be used, besides radiation introscopy, in the man radiation spectrometry, multiplex registration methods for charged particles, adamarov optical dispersion spectrometers and neutron-correlation spectrometry according to time-of-flight. In the latter case the matrix describes the slit arrangement in the neutron beam chopper. The given approach to optimization and the results obtained are also applicable for introscopes with spatial modulation of a signal

Additional details

Additional titles

Original title (Russian)
Оптимизация коллимирующих устройств гамма-камер с временной модуляцией сигнала

Publishing Information

Journal Title
Prib. Tekh. Ehksp.
Journal Issue
no.6
Series
Prib. Tekh. Ehksp.
ISSN
0032-8162

INIS

Country of Publication
Russian Federation
Country of Input or Organization
USSR
INIS RN
12575961
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
COLLIMATORS; GAMMA CAMERAS; MATRICES; MODULATION; OPTIMIZATION
Descriptors DEC
CAMERAS

Optional Information

Notes
For English translation see the journal Instruments and Experimental Techniques (USA).