Published February 15, 2006 | Version v1
Miscellaneous

CDTE Spectral Resolution Improvement through Rise-Time and Energy Based Correction Function

  • 1. 1Nuclear Research Center, Beer Sheva (Israel)
  • 2. Department of Biomedical Engineering, Ben Gurion University, Beer Sheva (Israel)

Description

To improve detection of low energy gamma ray in multi energy gamma ray measurement, it is essential to reduce the background signal. One way to achieve this goal is by reducing the measured amount of Comp ton's events produced by higher energy gamma that are part of the background in the lower gamma ray photo peaks. A portable detector that reduces background for low gamma ray (up to 1500 keV) by approximately 50%, thus shortening the detection time by 50%, has been developed. This paper describes the gamma detector elements, results and performances

Part of:
The 23. conference of the Nuclear Societies in Israel. Book of articles

Additional details

Publishing Information

Imprint Title
The 23. conference of the Nuclear Societies in Israel. Book of articles
Imprint Pagination
256 p.
Journal Page Range
p. 167-169
Report number
INIS-IL--012

Conference

Title
23. conference of the Nuclear Societies in Israel
Dates
15-16 Feb 2006
Place
Dead Sea (Israel)

Optional Information