CDTE Spectral Resolution Improvement through Rise-Time and Energy Based Correction Function
- 1. 1Nuclear Research Center, Beer Sheva (Israel)
- 2. Department of Biomedical Engineering, Ben Gurion University, Beer Sheva (Israel)
Description
To improve detection of low energy gamma ray in multi energy gamma ray measurement, it is essential to reduce the background signal. One way to achieve this goal is by reducing the measured amount of Comp ton's events produced by higher energy gamma that are part of the background in the lower gamma ray photo peaks. A portable detector that reduces background for low gamma ray (up to 1500 keV) by approximately 50%, thus shortening the detection time by 50%, has been developed. This paper describes the gamma detector elements, results and performances
Additional details
Publishing Information
- Imprint Title
- The 23. conference of the Nuclear Societies in Israel. Book of articles
- Imprint Pagination
- 256 p.
- Journal Page Range
- p. 167-169
- Report number
- INIS-IL--012
Conference
- Title
- 23. conference of the Nuclear Societies in Israel
- Dates
- 15-16 Feb 2006
- Place
- Dead Sea (Israel)
INIS
- Country of Publication
- Israel
- Country of Input or Organization
- Israel
- INIS RN
- 38093065
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CDTE SEMICONDUCTOR DETECTORS; MONTE CARLO METHOD; NEURAL NETWORKS; PULSE RISE TIME; RADIATION DETECTORS; RESOLUTION; SPECTRA UNFOLDING; SPECTRAL FUNCTIONS
- Descriptors DEC
- CALCULATION METHODS; DATA PROCESSING; FUNCTIONS; MEASURING INSTRUMENTS; PROCESSING; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; TIMING PROPERTIES