Published June 2007 | Version v1
Report

Carbon ion radiotherapy: Clinical aspects of dose reporting and treatment results at GSI

  • 1. Department of Radiation Oncology, University of Heidelberg (Germany)
  • 2. Division of Medical Physics in Radiation Oncology, German Cancer Research Center, Heidelberg (Germany)

Description

For carbon ion radiotherapy (RT), target volumes and organs at risk are defined according to the ICRU recommendations for conformal RT. Immobilization and precision of beam delivery are prerequisite. Treatment plan quality criteria used for the assessment of photon plans do also apply to carbon ion therapy plans. However, the reporting of physical doses alone is not sufficient, but biologically effective dose distributions and dose volume histogram (DVH) data have to be reported as well. Using active beam delivery techniques, the biological effective dose is not prescribed to a reference point but to each voxel. The α/β values used for biological treatment plan optimization need to be indicated as well. It is advisable to report the α/β values for the dose limiting toxicity and to optimize the plans according to this biological endpoint. The estimated α/β value for the specific tumour cell type and an additional treatment optimization for this endpoint might be reported as well in cases where α/β ratios for the dose-limiting toxicity and the specific tumour cell type are expected to differ substantially. The data reported should be compatible with the existing ICRU recommendations and should be sufficient to allow for clinical validation of the biological model used for biological plan optimization. Correction of minor systematic deviances of e.g. less than 5% should be enabled. Regular updates of the treatment planning software due to improvements/changes of the underlying beam model are needed in the future. To facilitate retrospective correlation of the collected clinical data with the model estimations and intercomparison of clinical data between different particle centres, the applied planning software version needs to be reported for each patient in the treatment protocol. (author)

Part of:
Dose reporting in ion beam therapy. Proceedings of a meeting

Additional details

Publishing Information

ISBN
978-92-0-105807-2
Imprint Title
Dose reporting in ion beam therapy. Proceedings of a meeting
Imprint Pagination
189 p.
Journal Page Range
p. 131-137
ISSN
1011-4289
Report number
IAEA-TECDOC--1560

Conference

Title
Meeting on dose reporting in ion beam therapy
Dates
18-20 Mar 2006
Place
Colombus, OH (United States)

INIS

Country of Publication
International Atomic Energy Agency (IAEA)
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
38079793
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ACCURACY; BIOLOGICAL MODELS; CARBON IONS; COMPUTER CODES; CORRECTIONS; ION BEAMS; OPTIMIZATION; ORGANS; PATIENTS; PHOTON BEAMS; PLANNING; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; RECOMMENDATIONS; TOXICITY; VALIDATION
Descriptors DEC
BEAMS; BODY; CHARGED PARTICLES; DOSES; IONS; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; TESTING; THERAPY

Optional Information

Notes
16 refs