Published October 2010 | Version v1
Journal article

Impurity temperature correction factors for the transmission grating spectrometer in the TJ-II stellarator

  • 1. Laboratorio Nacional de Fusion, Association Euratom-Ciemat, E-28040 Madrid (Spain)

Description

Impurity ion temperature and velocity profiles are obtained across plasmas in the TJ-II stellarator by performing charge-exchange recombination spectroscopy with a diagnostic neutral beam injector. For this, a tridirectional (toroidal plus two poloidal opposing views) multichannel spectroscopic diagnostic, incorporating 12-way fiber arrays, a compact f/1.8 spectrograph, and a back-illuminated CCD, permits Doppler line shifts and widths (of the C VI line at 529.05 nm) to be determined with 1-2 cm spatial resolution. For good photon counting statistics under Li-coated wall conditions, 600 μm diameter fibers collect and transmit light to curved 100 μm wide input slits. When calibrated with a neon pencil lamp this entrance slit width results in a non-Gaussian instrumental function that, if not handled correctly, can result in systematically underestimated impurity temperatures. Here we develop and present correction factors for this effect for a range of conditions.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
81
Journal Issue
10
Journal Page Range
p. 10D705-10D705.3
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2010 American Institute of Physics