Phantom study of the impact of reconstruction parameters on the detection of mini- and micro-volume lesions with a low-dose PET/CT acquisition protocol
Creators
- 1. Department of Nuclear Medicine and PET/CT Centre, Santa Maria della Misericordia Hospital, Via Tre Martiri 140, 45100 Rovigo (Italy)
- 2. Department of Medical Physics, Santa Maria della Misericordia Hospital, Via Tre Martiri 140, 45100 Rovigo (Italy)
Description
Purpose: Every PET scanner suffers of the partial volume effect (PVE), that is a loss of contrast in small lesions causing a worsening in standardized uptake value (SUV) accuracy, that is critical if quantitative PET/CT imaging is used for diagnosis and therapy. Methods: In order to quantify PVE and optimize our clinical protocols to minimize this effect in a last generation PET/CT scanner, we utilized a cylindrical phantom equipped with ten mini- and micro-volume hollow spheres. The lesion detectability and the SUV accuracy were evaluated at a fixed spheres to background intrinsic contrast (activity concentration ratio 8:1) but in different scan conditions: (a) acquisition modality (3D vs. 2D), (b) number of subset per iteration, (c) type of post-reconstruction filter and (d) activity concentration (i.e. total counts). Also the effect of different absorber thickness was evaluated. Results: Small lesion detectability resulted better in images acquired in 3D mode rather than 2D, mainly because of the lower noise produced by the fully-3D algorithm. The number of reconstruction iterations and the post-processing filter used affected both the contrast underestimation and the spatial resolution. Decreasing the 18F activity injected according to the low-dose protocol, the small lesions could be distinguished from the background down to a diameter of 6.2 mm and the SUV accuracy did not deteriorate. Adding absorber thickness around the phantom, the image noise slightly increased while SUV accuracy did not change. Conclusions: The hybrid PET/CT scanner we evaluated showed good performances, mainly in 3D acquisition modality. The phantom measurements showed that the most appropriate reconstruction protocol derived from a compromise between the contrast accuracy and the noise variance in PET images. The low-dose protocol clinically used demonstrated no loss in SUV accuracy and an adequate lesion detectability for lesions down to 6.2 mm in diameter.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.ejrad.2012.05.001Additional details
Identifiers
- DOI
- 10.1016/j.ejrad.2012.05.001;
- PII
- S0720-048X(12)00209-4;
Publishing Information
- Journal Title
- European Journal of Radiology
- Journal Volume
- 81
- Journal Issue
- 11
- Journal Page Range
- p. 3363-3370
- ISSN
- 0720-048X
- CODEN
- EJRADR
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 45027676
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALGORITHMS; CAT SCANNING; CONCENTRATION RATIO; DIAGNOSIS; FLUORINE 18; IMAGES; PERFORMANCE; PHANTOMS; POSITRON COMPUTED TOMOGRAPHY; SPATIAL RESOLUTION; UPTAKE
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-PLUS DECAY RADIOISOTOPES; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DIMENSIONLESS NUMBERS; EMISSION COMPUTED TOMOGRAPHY; FLUORINE ISOTOPES; HOURS LIVING RADIOISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; LIGHT NUCLEI; MATHEMATICAL LOGIC; MOCKUP; NANOSECONDS LIVING RADIOISOTOPES; NUCLEI; ODD-ODD NUCLEI; RADIOISOTOPES; RESOLUTION; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Copyright
- Copyright (c) 2012 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.