Qualification Tests for the New Air Sampling System at the 296-Z-1 Stack

Description

This report documents tests performed by Pacific Northwest National Laboratory to verify that the new air monitoring system for the 296-Z-1 ventilation exhaust stack meets the applicable regulatory criteria regarding the placement of the air sampling probe, sample transport, and stack flow measurement accuracy. These criteria ensure that the contaminants in the stack are well mixed with the airflow at the location of the probe so that the collected sample represents the whole. The sequence of tests addresses the acceptability of the flow angle relative to the probe uniformity of air velocity and gaseous and particle tracers in the cross section of the stack delivery of the sample from the sampler nozzle to the collection filter. The tests conducted on the air monitoring system demonstrated that the location for the air-sampling probe meets all performance criteria for air sampling systems at nuclear facilities. The performance criterion for particle transport was also met. All tests were successful, and all acceptance criteria were met

Availability note (English)

Available from PURL: https://www.osti.gov/servlets/purl/15010055-HMIvzv/native/

Additional details

Publishing Information

Imprint Pagination
[vp.]
Report number
PNNL--14057

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
36026654
Subject category
S11: NUCLEAR FUEL CYCLE AND FUEL MATERIALS;
Resource subtype / Literary indicator
Non-conventional Literature
Descriptors DEI
ACCURACY; AIR; CROSS SECTIONS; MONITORING; NOZZLES; NUCLEAR FACILITIES; PERFORMANCE; PROBES; SAMPLERS; SAMPLING; TRANSPORT; VELOCITY; VENTILATION
Descriptors DEC
EQUIPMENT; FLUIDS; GASES

Optional Information

Contract/Grant/Project number
820201000; AC06-76RL01830
Funding organization
US Department of Energy (United States)