Published April 11, 2002
| Version v1
Journal article
Search for an optimum time response of spark counters
Description
A spark counter of the type developed by Pestov has been tested with the aim of searching for an optimum time response function, changing voltage, content of noble and quencher gases, pressure and energy-loss. Replacing the usual argon by neon has brought an improvement of the resolution and a significant reduction of tails in the time response function. It has been proven that a counter as long as 90 cm can deliver, using neon gas mixture, a time resolution σ<60 ps with about 1% absolute tail and an efficiency of about 90%
Additional details
Identifiers
- PII
- S0168900201016229;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 482
- Journal Issue
- 1-2
- Journal Page Range
- p. 179-191
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34002539
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ENERGY LOSSES; NEON; OPTIMIZATION; RARE GASES; SPARK COUNTERS; TIME RESOLUTION; TIME-OF-FLIGHT SPECTROMETERS
- Descriptors DEC
- ELEMENTS; FLUIDS; GASES; LOSSES; MEASURING INSTRUMENTS; NONMETALS; RADIATION DETECTORS; RARE GASES; RESOLUTION; SPECTROMETERS; TIMING PROPERTIES
Optional Information
- Copyright
- Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.