Monte Carlo simulation of the Tomotherapy treatment unit in the static mode using MC HAMMER, a Monte Carlo tool dedicated to Tomotherapy
- 1. Department of Radiotherapy, St-Luc University Hospital, 10 av. Hippocrate, 1200 Brussels (Belgium)
- 2. Tomotherapy Inc., 1240 Deming Way, Madison, Wisconsin 52717 (United States)
- 3. Facultat de Fisica (ECM), Universitat de Barcelona, Societat Catalan de Fisica (IEC), Diagonal 647, 08028 Barcelona (Spain)
Description
Helical tomotherapy (HT) is designed to deliver highly modulated IMRT treatments. The concept of HT provides new challenges in MC simulation, because simultaneous movement of the gantry, the couch and the multi-leaf collimator (MLC) must be simulated accurately. However, before accounting for gantry, couch movement and multileaf collimator configurations, high accuracy must be achieved while simulating open static fields (1 x 40, 2.5 x 40 and 5 x 40 cm2). This is performed using MC HAMMER, which is a graphical user interface allowing MC simulation using PENELOPE for various configurations of HT. Since the geometry of the different elements and materials involved in the beam generation are precisely known and defined, the only parameters that need to be tuned on are therefore electron source spot size and electron energy. Beyond the build up region, good agreement (2%/1mm) is achieved for all the field sizes between measurements (ion chamber) and simulations with an electron source energy set to 5.5 MeV. The electron source spot size is modelled as a gaussian distribution with full width half maximum equal to 1.4 mm. This value was chosen to match measured and calculated penumbras in the longitudinal direction
Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Physics. Conference Series (Online)
- Journal Volume
- 74
- Journal Issue
- 1
- Journal Page Range
- p. 021019
- ISSN
- 1742-6596
Conference
- Title
- 1. European workshop on Monte Carlo treatment planning
- Dates
- 22-25 Oct 2006
- Place
- Gent (Belgium)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38080224
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCOUNTING; ACCURACY; COMPUTERIZED SIMULATION; COMPUTERIZED TOMOGRAPHY; ELECTRONS; GAUSS FUNCTION; M CODES; MEV RANGE; MONTE CARLO METHOD; P CODES; RADIOTHERAPY
- Descriptors DEC
- CALCULATION METHODS; COMPUTER CODES; DIAGNOSTIC TECHNIQUES; ELEMENTARY PARTICLES; ENERGY RANGE; FERMIONS; FUNCTIONS; LEPTONS; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; SIMULATION; THERAPY; TOMOGRAPHY