Optical fibre dosemeter systems for clinical applications based on radioluminescence and optically stimulated luminescence from Al2O3:C

  • 1. Radiation Research Dept., Risoe National Laboratory, DK-4000 Roskilde (Denmark)

Description

Optical fibre dosemeter systems based on radioluminescence and optically stimulated luminescence (OSL) from carbon-doped aluminium oxide (Al2O3:C) crystals were developed for in vivo real-time dose rate and absorbed dose measurements in radiotherapy and mammography. A technique was also developed for making ultra-small dosemeter probes that can easily be placed inside patients in radiation treatment. These probes have shown excellent properties in both head and neck intensity-modulated radiation therapy treatment and in mammography. The dose-response of the OSL signal for the new optical fibre dosemeter system showed a repeatability within 0.15% at a dose level of 60 mGy when integrated over 100 s. The temperature dependence in the range 0-45 deg. C showed a reproducibility within 1.3% when the OSL signal was integrated over 100 s. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/nci506

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
120
Journal Issue
1-4
Journal Page Range
p. 28-32
ISSN
0144-8420

Conference

Title
14. International Conference on Solid State Dosimetry
Acronym
SSD14
Dates
27 Jun - 2 Jul 2004
Place
New Haven (United States)

Optional Information

Notes
11 refs