Evaluation of a deep learning-based pelvic synthetic CT generation technique for MRI-based prostate proton treatment planning

  • 1. Department of Radiation Oncology and Winship Cancer Institute, Emory University, Atlanta, GA 30322 (United States)

Description

The purpose of this work is to validate the application of a deep learning-based method for pelvic synthetic CT (sCT) generation that can be used for prostate proton beam therapy treatment planning. We propose to integrate dense block minimization into 3D cycle-consistent generative adversarial networks (cycleGAN) framework to effectively learn the nonlinear mapping between MRI and CT pairs. A cohort of 17 patients with co-registered CT and MR pairs were used to test the deep learning-based sCT generation method by leave-one-out cross-validation. Image quality between the sCT and CT images, gamma analysis passing rate, dose-volume metrics, distal range displacement, and the individual pencil beam Bragg peak shift between sCT- and CT-based proton plans were evaluated. The average mean absolute error (MAE) was 51.32  ±  16.91 HU. The relative differences of the statistics of the PTV dose-volume histogram (DVH) metrics in between sCT and CT were generally less than 1%. Mean values of dose difference, absolute dose difference (in percent of the prescribed dose) were  −0.07%  ±  0.07% and 0.23%  ±  0.08%. Mean gamma analysis pass rate of 1 mm/1%, 2 mm/2%, 3 mm/3% criteria with 10% dose threshold were 92.39%  ±  5.97%, 97.95%  ±  2.95% and 98.97%  ±  1.62% respectively. The median, mean and standard deviation of absolute maximum range differences were 0.09 cm and 0.23  ±  0.25 cm. The median and mean Bragg peak shifts among the 17 patients were 0.09 cm and 0.18  ±  0.07 cm. The image similarity, dosimetric and distal range agreement between sCT and original CT suggests the feasibility of further development of an MRI-only workflow for prostate proton radiotherapy. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1361-6560/ab41af

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
64
Journal Issue
20
Journal Page Range
[15 p.]
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
52004069
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BRAGG CURVE; IMAGE PROCESSING; LEARNING; NMR IMAGING; NONLINEAR PROBLEMS; PROSTATE; PROTON BEAMS; RADIATION DOSES; RADIOTHERAPY
Descriptors DEC
BEAMS; BODY; DIAGNOSTIC TECHNIQUES; DIAGRAMS; DOSES; GLANDS; INFORMATION; MALE GENITALS; MEDICINE; NUCLEAR MEDICINE; NUCLEON BEAMS; ORGANS; PARTICLE BEAMS; PROCESSING; RADIOLOGY; THERAPY