Published 2000
| Version v1
Miscellaneous
X-ray spectrometry with Si photodiodes
Description
This work envisages the development of a high resolution electromagnetic radiation spectrometry system, based on the use of low cost commercial silicon photodiodes (Siemens SFH00206 and Hamamatsu S2506-02). In order to improve the performance of these photodiodes for high resolution X-ray spectrometry, a cooling system using Peltier cells was developed to decrease the temperature of both photodiode and preamplifier's FET. The best values of the resolution for the 59.5 keV 241 Am γ-ray line, at 9 deg C, were found to be 2.10 keV and 1.82 keV for the Siemens SFH00206 and Hamamatsu S250602 photodiodes respectively. (author)
Availability note (English)
Available from the Library of the Brazilian Nuclear and Energy Research Institute (CNEN/IPEN), Sao Paulo, SPAdditional details
Additional titles
- Original title (Portuguese)
- Espectrometria de raios-X com diodos de Si
Publishing Information
- Imprint Pagination
- 81 p.
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 37079218
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Numerical Data, Thesis, Non-conventional Literature
- Descriptors DEI
- AMERICIUM 241; CALIBRATION STANDARDS; ENERGY RESOLUTION; EXPERIMENTAL DATA; PHOTODIODES; PREAMPLIFIERS; SEMICONDUCTOR DETECTORS; SILICON DIODES; TEMPERATURE DEPENDENCE; TEMPERATURE NOISE; X-RAY SPECTROSCOPY
- Descriptors DEC
- ACTINIDE NUCLEI; ALPHA DECAY RADIOISOTOPES; AMERICIUM ISOTOPES; AMPLIFIERS; DATA; ELECTRONIC EQUIPMENT; EQUIPMENT; HEAVY NUCLEI; INFORMATION; ISOTOPES; MEASURING INSTRUMENTS; NOISE; NUCLEI; NUMERICAL DATA; ODD-EVEN NUCLEI; RADIATION DETECTORS; RADIOISOTOPES; RESOLUTION; SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES; SPECTROSCOPY; SPONTANEOUS FISSION RADIOISOTOPES; STANDARDS; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- 30 refs., 56 figs., 9 tabs.