Published April 2016 | Version v1
Journal article

The effect of chest circumference index adjusting tube voltages technique on image quality and radiation dose at coronary CT angiography

  • 1. Department of Radiology, Lishui Central Hospital, Zhejiang Prorince, Lishui (China)

Description

Objective: To explore the effect of chest circumference index adjusting tube voltage technique on image quality and radiation dose at coronary CTA. Methods: One hundred and twenty consecutive patients [body mass index(BMI) < 25 kg/m2] with suspected coronary heart disease (CHD) undergoing coronary CT angiography were prospectively selected and divided into 2 groups at random: conventional group and low dose group. Low dose group was divided into the following three subgroups according to different chest circumferences: A group(< 80 cm, n = 16), B group(80 to 90 cm, n = 44) and C group (> 90 cm, n = 20). All patients were examined by coronary CTA. The patients in conventional group were performed using retrospective ECG-gating technology and reconstructed by filtered back projection algorithm. The tube voltage/tube current was 120 kV/1000 mAs. Prospective ECG-gating technology and iterative algorithm reconstruction were used in low dose group. The tube voltages/currents were 80 kV/150 mAs, 100 kV/150 mAs, 120 kV/150 mAs in A, B, C group, respectively. Image quality was assessed by subjective evaluation (image quality score) and objective evaluation (signal-to-noise ratio).The effective radiation dose was calculated. Analyses of the differences between groups were compared with image quality, radiation dose by single factor variance and Wilcoxon signed ranks test. Results: The image quality scores and signal-to-noise ratio of aorta were respectively (3.47 ± 0.38), (3.48 ± 0.27), (3.45 ± 0.32), (3.46 ± 0.29) and (15.5 ± 3.6), (15.8 ± 3.6), (15.8 ± 4.1), (16.2 ± 3.9) in conventional, A, B and C groups, there was no statistical difference between the four groups (P = 0.24, 0.43). The effective radiation dose of four groups were respectively (17.15 ± 3.25), (0.88 ± 0.02), (1.38 ± 0.05), (2.32 ± 0.04) mSv, the difference was statistically significant (P = 0.02). The effective radiation dose of A, B, C group was significantly lower than that in the conventional group. Conclusion: Chest circumference index adjusting tube voltages technology at coronary CT angiography can effectively reduce the effective radiation without compromise of image quality. (authors)

Additional details

Publishing Information

Journal Title
Chinese Journal of Radiology
Journal Volume
50
Journal Issue
4
Journal Page Range
p. 284-288
ISSN
1005-1201

Optional Information

Notes
8 figs., 2 tabs., 9 refs.; http://dx.doi.org/10.3760/cma.j.issn.1005-1201.2016.04.011