Published October 7, 2000 | Version v1
Journal article

Design considerations for transmission x-ray tubes operated at diagnostic energies

  • 1. Physics Department, University of Surrey, Guildford, Surrey (United Kingdom) and Departamento de Fisica, Universidade de Coimbra, Coimbra (Portugal)
  • 2. Physics Department, University of Surrey, Guildford, Surrey (GB)

Description

A low-energy expansion of the EGS4 Monte Carlo code system was used in studies that aimed to understand photon production in transmission target x-ray tubes. In each case, foil targets were modelled with incident electrons of a few tens of keV. The production of bremsstrahlung and fluorescence radiation at different depths was studied and the angular distribution of radiation analysed. The characteristics of x-ray spectra measured along the beam path, for targets of different materials, thicknesses and several beam energies have been determined and the relation between efficiency and the fluorescence-to-bremsstrahlung ratio is shown. (author)

Availability note (English)

Available online at the Web site for the Journal of Physics. D, Applied Physics (ISSN 1361-6463) http://www.iop.org/