Study of a multi-wire proportional chamber with a cathode strip and delay-line readout
- 1. School of Physics and State Key Laboratory of Nuclear Physics and Technology, Beijing Univ., Beijing (China)
Description
The design principle for a multi-wire proportional chamber with a cathode strip and delay-line readout is described. A prototype chamber of a size of 10 cm x 10 cm was made together with the readout electronics circuit. A very clean signal with very low background noise was obtained by applying a transformer between the delay-line and the pre-amplifier in order to match the resistance. Along the anode wire direction a position resolution of less than 0.5 mm was achieved with a 55Fe-5.9 keV X ray source. The simple structure, large effective area and high position resolution allow the application of a gas chamber of this kind to many purposes. (authors)
Additional details
Publishing Information
- Journal Title
- Chinese Physics. C, High Energy Physics and Nuclear Physics
- Journal Volume
- 33
- Journal Issue
- 5
- Journal Page Range
- p. 364-368
- ISSN
- 1674-1137
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 41125243
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BACKGROUND NOISE; CATHODES; DELAY CIRCUITS; MULTIWIRE PROPORTIONAL CHAMBERS; READOUT SYSTEMS; SPATIAL RESOLUTION; TRANSFORMERS
- Descriptors DEC
- ELECTRICAL EQUIPMENT; ELECTRODES; ELECTRONIC CIRCUITS; EQUIPMENT; MEASURING INSTRUMENTS; NOISE; PROPORTIONAL COUNTERS; RADIATION DETECTORS; RESOLUTION
Optional Information
- Notes
- 7 figs., 11 refs.