Description

Additional software and modification of the electronic circuit of the NOKIA LP-4900 analyzer allowing to perform the multiscaler mode without using the appropriate stadard LPH-4913 unit, which, as a rule, is not a part of the standard analyzer set, are described. Besides, modification gives the possibility, being absent before, to control the analyzer by means of external signals, as well as to control the external devices according to the analyzer commands. Application of the modified analyzer in the experiment on simulation of computer tomography is considered

Availability note (English)

MF available from INIS under the Report Number.

Files

18100809.pdf

Files (246.1 kB)

Name Size Download all
md5:fd6518827f3fa0936709aeaf966df3f7
246.1 kB Preview Download

System files (27.8 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Мультискейлерный режим в анализаторе NOKIA LP-4900 без использования стандартного блока LPH-4913

Publishing Information

Imprint Pagination
12 p.
Report number
IAE--4302/11

Optional Information

Notes
5 figs.