Published May 1981 | Version v1
Journal article

Spectral characterization of 4 MV Bremsstrahlung by attenuation analysis

  • 1. Joint Center for Radiation Therapy, Harvard Medical School, Boston, Massachusetts

Description

The quality of radiation for a high energy x-ray beam can be specified by its attenuation curve in a selected material. The inverse Laplace transform of the attenuation curve can be used as an approximate indication of the energy spectrum of the beam. Existing mathematical procedures for this purpose have been evaluated and were found to poorly represent measured transmission data for 4 MV x-rays from a linear accelerator. The transmission data between 1 and 0.002 could be fitted within the experimental uncertainty by expressing the logarithmic transmission as a second order polynomial of attenuator thickness. The inverse Laplace transform them becomes a Gaussian function of the attenuation coefficient. This new version of attenuation analysis provides a practical method for specification of the quality of the radiation in this energy range

Additional details

Publishing Information

Journal Title
Med. Phys.
Journal Volume
8
Journal Issue
3
Series
Med. Phys.
Journal Page Range
368-74
ISSN
0094-2405

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
14719858
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BREMSSTRAHLUNG; RADIATION QUALITY; RADIOTHERAPY; X-RAY SOURCES
Descriptors DEC
ELECTROMAGNETIC RADIATION; EQUIPMENT; MEDICINE; RADIATION SOURCES; RADIATIONS; THERAPY; X-RAY EQUIPMENT