Design (Ge-BGO+NaI(Tl)) of Anti-Compton Spectrometer

  • 1. Joint Inst. for Nuclear Research, Dubna (Russian Federation). Lab. of Nuclear Problems
  • 2. Ceskoslovenska Akademie Ved, Rez (Czech Republic). Ustav Jaderne Fyziky
  • 3. Monokrystaly, Turnov (Czech Republic)

Description

The design of an asymmetric anti-Compton shield for BGO and NaI(Tl) crystals is described. Its dimensions have been optimized by means of Monte Carlo calculations over the range of incident gamma-ray energies between 0.2 MeV and 3.5 MeV. Monte Carlo method takes into account photoelectric absorption of γ-rays, Compton scattering and pair production. The suppression factor of the continuous 60 Co spectrum for the BGO and NaI(Tl) crystals of different dimensions is calculated for anti-Compton shield. (author). 25 refs., 9 figs., 2 tabs

Availability note (English)

MF available from INIS under the Report Number.

Files

26011620.pdf

Files (407.4 kB)

Name Size Download all
md5:aa0ffb1cdfce860023ed1c0494b4c414
407.4 kB Preview Download

System files (46.9 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Расчет антикомптоновского (Ге-BGO+NaI(Тл))-спеcтрометра. (Проект прибора)

Publishing Information

Imprint Pagination
22 p.
Report number
JINR-R--13-94-277