Published October 2017
| Version v1
Journal article
Design of an on-line monitoring system for radioactive wastewater
- 1. Shenyang Institute of Engineering, Liaoning (China)
Description
An on-line monitoring system for radioactive wastewater was designed to discriminate the type and concentration of the radionuclides discharged from nuclear facilities. An HPGe semiconductor was used as the detector in the system for continuous monitoring by pumping wastewater. The minimum detectable activity for 137Cs was 0.4 Bq L-1 after 10 min of measuring wastewater with the system. The system can detect excessive radioactivity in the wastewater and quickly and effectively alert personnel. Based on the experimental measurements and the Monte Carlo simulation, the detection efficiency of the system was calibrated, and an efficiency curve was determined for the energy range from 50 to 2754 keV. (author)
Additional details
Publishing Information
- Journal Title
- Journal of Radioanalytical and Nuclear Chemistry
- Journal Volume
- 314
- Journal Issue
- 1
- Journal Page Range
- p. 215-220
- ISSN
- 0236-5731
- CODEN
- JRNCDM
INIS
- Country of Publication
- Hungary
- Country of Input or Organization
- Hungary
- INIS RN
- 48085628
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CESIUM 137; HIGH-PURITY GE DETECTORS; KEV RANGE; MONITORING; NUCLEAR FACILITIES; ON-LINE MEASUREMENT SYSTEMS; PERSONNEL; PUMPING; RADIOACTIVE WASTE MANAGEMENT; RADIOACTIVITY; WASTE WATER
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CESIUM ISOTOPES; ENERGY RANGE; GE SEMICONDUCTOR DETECTORS; HYDROGEN COMPOUNDS; INTERMEDIATE MASS NUCLEI; ISOTOPES; LIQUID WASTES; MANAGEMENT; MEASURING INSTRUMENTS; NUCLEI; ODD-EVEN NUCLEI; ON-LINE SYSTEMS; OXYGEN COMPOUNDS; RADIATION DETECTORS; RADIOISOTOPES; SEMICONDUCTOR DETECTORS; WASTE MANAGEMENT; WASTES; WATER; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- 15 refs.