Published September 1, 2010 | Version v1
Journal article

High-Frequency Jet Ventilation for Complete Target Immobilization and Reduction of Planning Target Volume in Stereotactic High Single-Dose Irradiation of Stage I Non-Small Cell Lung Cancer and Lung Metastases

  • 1. Department of Radiotherapy, St. Marien-Krankenhaus, Siegen (Germany)
  • 2. Department of Anaesthesiology, St. Marien-Krankenhaus, Siegen (Germany)
  • 3. Institute of Medical Biometry and Epidemiology, University of Marburg (Germany)

Description

Purpose: To demonstrate the feasibility of complete target immobilization by means of high-frequency jet ventilation (HFJV); and to show that the saving of planning target volume (PTV) on the stereotactic body radiation therapy (SBRT) under HFJV, compared with SBRT with respiratory motion, can be predicted with reliable accuracy by computed tomography (CT) scans at peak inspiration phase. Methods and Materials: A comparison regarding different methods for defining the PTV was carried out in 22 patients with tumors that clearly moved with respiration. A movement span of the gross tumor volume (GTV) was defined by fusing respiration-correlated CT scans. The PTV enclosed the GTV positions with a safety margin throughout the breathing cycle. To create a PTV from CT scans acquired under HFJV, the same margins were drawn around the immobilized target. In addition, peak inspiration phase CT images (PIP-CTs) were used to approximate a target immobilized by HFJV. Results: The resulting HFJV-PTVs were between 11.6% and 45.4% smaller than the baseline values calculated as respiration-correlated CT-PTVs (median volume reduction, 25.4%). Tentative planning by means of PIP-CT PTVs predicted that in 19 of 22 patients, use of HFJV would lead to a reduction in volume of ≥20%. Using this threshold yielded a positive predictive value of 0.89, as well as a sensitivity of 0.94 and a specificity of 0.5. Conclusions: In all patients, SBRT under HFJV provided a reliable immobilization of the GTVs and achieved a reduction in PTVs, regardless of patient compliance. Tentative planning facilitated the selection of patients who could better undergo radiation in respiratory standstill, both with greater accuracy and lung protection.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.ijrobp.2009.07.1678

Additional details

Identifiers

DOI
10.1016/j.ijrobp.2009.07.1678;
PII
S0360-3016(09)02717-5;

Publishing Information

Journal Title
International Journal of Radiation Oncology, Biology and Physics
Journal Volume
78
Journal Issue
1
Journal Page Range
p. 136-142
ISSN
0360-3016
CODEN
IOBPD3

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
42031671
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
CAT SCANNING; LUNGS; METASTASES; NEOPLASMS; PLANNING; RADIOTHERAPY; REDUCTION; VENTILATION
Descriptors DEC
BODY; CHEMICAL REACTIONS; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DISEASES; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; RESPIRATORY SYSTEM; THERAPY; TOMOGRAPHY

Optional Information

Copyright
Copyright (c) 2010 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.