Description

A thermoluminescent dosimeter (TLD) badge has been developed to measure beta and gamma doses for routine personal dosimetry. The badge uses a card with four elements of LiF:Mg,Ti enriched in 7Li and sandwiched between layers of TeflonTM. The Element 1 filter is 1000mgcm-2 of low-Z material, the Element 2 filter is 464mgcm-2 of tin plus 536mgcm-2 of low-Z material, the Element 3 filter is a 5-7mgcm-2 Mylar window, and the Element 4 filter is 64mgcm-2 of low-Z material. The filters for Elements 2 and 4 were optimized by testing and by application of radiation physics principles and calculation. Elements 1 and 2 are used to measure the response to photons, and from these results, Hp(10), Hp(0.07) for photons, and the photon contribution to the signals for Elements 3 and 4 are obtained. The photon contribution to the signals from Elements 3 and 4 is subtracted from their total signals to obtain two signals due to betas. These are used to derive a corrected value for Hp(0.07) for betas. The badge has been in use since early 1999, and provides accurate results in a nuclear power station environment

Availability note (English)

Available from http://dx.doi.org/10.1016/j.radmeas.2007.11.053

Additional details

Identifiers

DOI
10.1016/j.radmeas.2007.11.053;
PII
S1350-4487(07)00503-3;

Publishing Information

Journal Title
Radiation Measurements
Journal Volume
43
Journal Issue
2-6
Journal Page Range
p. 525-532
ISSN
1350-4487
CODEN
RMEAEP

Conference

Title
15. solid state dosimetry conference
Acronym
SSD15
Dates
8-13 Jul 2007
Place
Delft (Netherlands)

Optional Information

Copyright
Copyright (c) 2007 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.