The device for the poloidal profile measurement of H sub(α)-line emission by photodiode and its calibration

  • 1. Japan Atomic Energy Research Inst., Naka, Ibaraki. Naka Fusion Research Establishment

Description

The device for the poloidal profile measurement of H sub(α)-line emission has been equipped by photodiode (S1225-5BQ by HAMAMATSU PHOTONICS K.K.) and operational amplifier. The absolute efficiency was calibrated by using He-Ne laser. The device is constructed by 44 channels. The fast type of 8 channels is for the monitor of pellet abration profile. The slow type of 36 channels is for the poloidal profile measurement of H sub(α)-line emission from JFT-2M plasma. The rise time of the fast type and the slow type is about 2.8 μsec and about 350 μsec, respectively. The absolute efficiency of the fast type and the slow type is 72.7 V/mW and 18.2 V/μW, respectively. (author)