Cardiac CT: are we underestimating the dose? A radiation dose study utilizing the 2007 ICRP tissue weighting factors and a cardiac specific scan volume
- 1. Plymouth Hospitals NHS Trust, Derriford Hospital, Plymouth, Devon (United Kingdom)
- 2. Peninsula College of Medicine and Dentistry, University of Plymouth, Devon (United Kingdom)
Description
Aim: To calculate the effective dose from cardiac multidetector computed tomography (MDCT) using a computer-based model utilizing the latest International Commission on Radiation Protection (ICRP) 103 tissue-weighting factors (2007), to compare this dose with those calculated with previously published chest conversion factors and to produce a conversion factor specific for cardiac MDCT. Materials and methods: An observational study of 152 patients attending for cardiac MDCT as part of their usual clinical care in a university teaching hospital. The dose for each examination was calculated using the computer-based anthropomorphic ImPACT model (the imaging performance assessment of CT scanners) and this was compared with the dose derived from the dose-length product (DLP) and a chest conversion factor. Results: The median effective dose calculated using the ImPACT calculator (4.5 mSv) was significantly higher than the doses calculated with the chest conversion factors (2.2-3 mSv). Conclusion: The use of chest conversion factors significantly underestimates the effective dose when compared to the dose calculated using the ImPACT calculator. A conversion factor of 0.028 would give a better estimation of the effective dose from prospectively gated cardiac MDCT.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.crad.2010.08.001Additional details
Identifiers
- DOI
- 10.1016/j.crad.2010.08.001;
- PII
- S0009-9260(10)00303-X;
Publishing Information
- Journal Title
- Clinical Radiology
- Journal Volume
- 65
- Journal Issue
- 12
- Journal Page Range
- p. 1013-1017
- ISSN
- 0009-9260
- CODEN
- CLRAAG
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 42080828
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CALCULATORS; CHEST; COMPUTERIZED TOMOGRAPHY; ICRP; PERFORMANCE; RADIATION DOSES; RADIATION PROTECTION
- Descriptors DEC
- BODY; COMPUTERS; DIAGNOSTIC TECHNIQUES; DIGITAL COMPUTERS; DOSES; INTERNATIONAL ORGANIZATIONS; TOMOGRAPHY
Optional Information
- Copyright
- Copyright (c) 2010 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.