Published December 1997 | Version v1
Report

Quality assurance regime for the fabrication and calibration of ambient dose rate measuring instruments with proportional counters

  • 1. Thermo Instrument Systems GmbH, Eberline Strahlen und Umweltmesstechnik, Erlangen (Germany)

Description

The proportional detectors FHZ 600A and FHZ 601A serve as examples to illustrate the quality assurance regime applied in fabrication and operation of the detector. Calibration is done by a licensed establishment connected to the standard 137Cs calibration source at PTB. In addition to the basic calibration, the intrinsic selfdosing effect of the device and its response to cosmic radiation are known and can be taken into account in interpretations of measured results. The individual calibration of the intelligent probes of type FHZ 601A achieves a yet better measuring accuracy at higher dose rates, as compared to the basic calibration and performance of the FHZ 600A probes, as was shown by a type approval test. The paper also shows that control measurements and ''intelligent'' measuring systems with integrated checking routines improve the measuring reliability. (orig./CB)

Abstract (German)

Am Beispiel der Proportionaldetektoren FHZ 600A und FHZ 601A werden die Qualitaetssicherungsmassnahmen bei der Herstellung und beim Betrieb des Detektors erlaeutert. Die Kalibrierung erfolgt an einer zur Eichung von Strahlenmessgeraeten zugelassenen und an das Normal der PTB angeschlossenen 137Cs-Kalibriereinrichtung. Neben dieser Grundkalibrierung ist zur Messung der natuerlichen Umgebungsstrahlung der Eigennulleffekt und das Ansprechvermoegen gegenueber der kosmischen Strahlung bekannt und kann zur Interpretation der Messdaten beruecksichtigt werden. Die individuelle Kalibrierung der intelligenten Sonden vom Type FHZ 601A bringt im hoeheren Dosisleistungsbereich nochmals Genauigkeitsvorteile gegenueber einer Einstellung mit mittleren Kalibrierwerten beim FHZ 600A, die im Rahmen einer Typpruefung ermittelt wurden. Ausserdem wird gezeigt, wie staendige Kontrollmessungen und 'intelligente' Messsysteme mit integrierten Pruefroutinen die Messsicherheit verbessern koennen. (orig.)
Part of:
110. PTB seminar: Dose rate measurements of ionizing radiation in the range of natural ambient radiation. Proceedings

Additional details

Additional titles

Original title (German)
Qualitaetssicherung bei der Herstellung und Kalibrierung von Dosisleistungsmessern mit Proportionalzaehlrohren

Publishing Information

ISBN
3-89701-054-2
Imprint Title
110. PTB seminar: Dose rate measurements of ionizing radiation in the range of natural ambient radiation. Proceedings
Imprint Pagination
192 p.
Journal Page Range
p. 33-41
ISSN
0172-7095
Report number
PTB-Dos--26

Conference

Title
Dose rate measurements of ionizing radiation in the range of natural ambient radiation
Original Conference Title
110. PTB-Seminar: Ortsdosismessungen Ionisierender Strahlung im Bereich der Natuerlichen Umgebungsstrahlung
Acronym
110. PTB seminar
Dates
30 Nov - 1 Dec 1993
Place
Braunschweig (Germany)

INIS

Country of Publication
Germany
Country of Input or Organization
Germany
INIS RN
30005847
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
ACCURACY; BACKGROUND RADIATION; CALIBRATION; DOSE EQUIVALENTS; DOSIMETRY; EFFICIENCY; PROPORTIONAL COUNTERS; QUALITY ASSURANCE; RADIATION MONITORING
Descriptors DEC
MEASURING INSTRUMENTS; MONITORING; RADIATION DETECTORS; RADIATIONS

Optional Information

Notes
Imprint:110. PTB-Seminar: Ortsdosisleistungsmessungen ionisierender Strahlung im Bereich der natuerlichen Umgebungsstrahlung. Vortraege