Published September 21, 2007 | Version v1
Journal article

Photon and fast neutron dosimetry using aluminium oxide thermoluminescence dosemeters in a pool-type research reactor

  • 1. Instituto Tecnologico e Nuclear, Estrada Nacional 10, 2683-953 Sacavem (Portugal)
  • 2. Hospital de Santa Maria, Av. Prof. Egas Moniz, 1649-035 Lisbon (Portugal)
  • 3. Institute of Isotopes and Surface Chemistry, Budapest, P.O. Box 77, H-1525 (Hungary)

Description

Al2O3:Mg,Y thermoluminescence (TL) dosemeters were used to measure photon and fast neutron doses in the mixed radiation field of the Portuguese Research Reactor. The dosemeters were irradiated in core positions under a photon dose rate of the order of 104 Gy/h and a fast neutron flux in the range of 109-1011 n/cm2/s. In order to evaluate the ability of the TL dosemeters for mixed field dosimetry at the research reactor, the measurements were compared with results obtained via conventional methods. The agreement between the different methods is better than 13% for the determination of photon doses and within 5% for the determination of neutron fluxes in mixed fields

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2007.05.153

Additional details

Identifiers

DOI
10.1016/j.nima.2007.05.153;
PII
S0168-9002(07)01037-6;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
580
Journal Issue
1
Journal Page Range
p. 310-313
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
10. international symposium on radiation physics
Acronym
ISRP 10
Dates
17-22 Sep 2006
Place
Coimbra (Portugal)

Optional Information

Copyright
Copyright (c) 2007 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.