Published June 2015 | Version v1
Journal article

SU-E-P-49: Evaluation of Image Quality and Radiation Dose of Various Unenhanced Head CT Protocols

  • 1. Beaumont Health System, Royal Oak, MI (United States)

Description

Purpose: To evaluate the diagnostic value of various unenhanced head CT protocols and predicate acceptable radiation dose level for head CT exam. Methods: Our retrospective analysis included 3 groups, 20 patients per group, who underwent clinical routine unenhanced adult head CT examination. All exams were performed axially with 120 kVp. Three protocols, 380 mAs without iterative reconstruction and automAs, 340 mAs with iterative reconstruction without automAs, 340 mAs with iterative reconstruction and automAs, were applied on each group patients respectively. The images were reconstructed with H30, J30 for brain window and H60, J70 for bone window. Images acquired with three protocols were randomized and blindly reviewed by three radiologists. A 5 point scale was used to rate each exam The percentage of exam score above 3 and average scores of each protocol were calculated for each reviewer and tissue types. Results: For protocols without automAs, the average scores of bone window with iterative reconstruction were higher than those without iterative reconstruction for each reviewer although the radiation dose was 10 percentage lower. 100 percentage exams were scored 3 or higher and the average scores were above 4 for both brain and bone reconstructions. The CTDIvols are 64.4 and 57.8 mGy of 380 and 340 mAs, respectively. With automAs, the radiation dose varied with head size, resulting in 47.5 mGy average CTDIvol between 39.5 and 56.5 mGy. 93 and 98 percentage exams were scored great than 3 for brain and bone windows, respectively. The diagnostic confidence level and image quality of exams with AutomAs were less than those without AutomAs for each reviewer. Conclusion: According to these results, the mAs was reduced to 300 with automAs OFF for head CT exam. The radiation dose was 20 percentage lower than the original protocol and the CTDIvol was reduced to 51.2 mGy

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
42
Journal Issue
6
Journal Page Range
vp.
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
47056682
Subject category
S60: APPLIED LIFE SCIENCES;
Descriptors DEI
ANIMAL TISSUES; BRAIN; COMPUTERIZED TOMOGRAPHY; HEAD; IMAGES; ITERATIVE METHODS; PATIENTS; RADIATION DOSES; REVIEWS; SKELETON
Descriptors DEC
BODY; CALCULATION METHODS; CENTRAL NERVOUS SYSTEM; DIAGNOSTIC TECHNIQUES; DOCUMENT TYPES; DOSES; NERVOUS SYSTEM; ORGANS; TOMOGRAPHY

Optional Information

Notes
(c) 2015 American Association of Physicists in Medicine