Description
With the technology of coincidence method, transit time spreads (TTSs) of PMTs are measured with Cherenkov light on the photocathode widow of photomultiplier tube (PMT) produced by cascade γ radiation source, 60Co. TTSs with single and multi-photoelectron which are analyzed by convolution of Poisson and Gaussian are obtained. The study to XP2020 and XP2020Q PMTs shows that TTSs are consistent with data supplied by Philips. TTSs are inversely proportional to the square roots of number of photoelectrons. The method is feasible when single and multi-photoelectron can be distinguished. (authors)
Additional details
Publishing Information
- Journal Title
- Nuclear Science and Techniques
- Journal Volume
- 20
- Journal Issue
- 6
- Journal Page Range
- p. 344-348
- ISSN
- 1001-8042
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 42093370
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CHERENKOV RADIATION; COBALT 60; COINCIDENCE METHODS; GAMMA RADIATION; PHOTOCATHODES; PHOTOMULTIPLIERS; ROOTS; TUBES
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CATHODES; COBALT ISOTOPES; COUNTING TECHNIQUES; ELECTRODES; ELECTROMAGNETIC RADIATION; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; IONIZING RADIATIONS; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MINUTES LIVING RADIOISOTOPES; NUCLEI; ODD-ODD NUCLEI; PHOTOTUBES; RADIATIONS; RADIOISOTOPES; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- 5 figs., 3 tabs., 9 refs.