Published February 1977 | Version v1
Journal article

Reduction of the low energy background of a phoswich detector

  • 1. UKAEA Research Group, Harwell. Atomic Energy Research Establishment

Description

Modifications to a phoswich detector and its attendant electronics to suppress contributions from dark current which can effect a worthwhile reduction to the background contribution in the 12 to 40 keV range are reported. The comparative background spectra of a 20 cm diameter phoswich with (a) a single 18 cm diameter photomultiplier and (b) three 7.5 cm diameter photomultipliers and multichannel analyser triggered by the coincident output from two tubes, are shown graphically. Reasons for the improved performance with the second arrangement are considered. (U.K.)

Additional details

Publishing Information

Journal Title
Health Phys.
Journal Volume
32
Journal Issue
2
Series
Health Phys.
Journal Page Range
113-116

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
8313433
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
BACKGROUND NOISE; ELECTRONIC CIRCUITS; ENERGY SPECTRA; KEV RANGE 10-100; MULTI-CHANNEL ANALYZERS; PHOTOMULTIPLIERS
Descriptors DEC
ELECTRONIC EQUIPMENT; ENERGY RANGE; KEV RANGE; NOISE; PHOTOTUBES; PULSE ANALYZERS; SPECTRA