Published January 1986 | Version v1
Journal article

Micro-determination of ytterbium with electrothermal atomic absorption spectrometry

  • 1. Academia Sinica, BJ (China). Inst. of Chemistry

Description

This communication reports the use of a pyrolytic graphite coated tube, lined with tantalum-tungsten, and a local made atomic absorption spectrometer (Model WFD-Y3) for the determination of small amount Yb in pure Y2O3 and mixed rare earth oxides. It is found that the method proposed is sensitive, reproducible and simple in manipulation. Even as low as 0.2 μg Yb in one gram sample (n x 10-7) can be determined directly without pre-concentration. It is found experimentally that the optimum condition for drying is at 150 deg C. for 20 sec, ashing at 1000 deg C. for 20 sec and atomization at 2770 deg C. for 12 sec. Within the range 1.0-18ng Yb/ml the calibration curve of Yb is linear. Before injecting into the tube, the acidity of the sample solution should be ajusted to 0.1 to 2 M with nitric or hydrochloric acid. For 5ng Yb/ml, Al(III), Ca(II) and La(III) interference, when their amount present is 50 μg/ml or more. On the other hand, Cu(II), Fe(III), Mg(II), K(I) and Y(III) in amount up to 1 mg/ml do not interfere

Additional details

Publishing Information

Journal Title
Anal. Chem. (Changchung)
Journal Volume
14
Journal Issue
1
Series
Anal. Chem. (Changchung).
Journal Page Range
8-12
CODEN
FHHHD