Published 2018 | Version v1
Book

Application of simulation technique for assessment of radioactivity in sample of irregular geometry

  • 1. Radiation Safety Systems Division, Bhabha Atomic Research Centre, Mumbai (India)

Description

Irregular shaped radioactive samples are encountered during few occasions. Radioactivity assessment of such samples is essential for proper disposal of the material. In this present work a simulation technique is used to calibrate 3 inches x 3 inches NaI(Tl) detector for non-conventional geometries or irregular shape samples. The methodology consists of development of efficiency parameter from the simulated spectrum using detector and collimator model. The efficiency evaluated will be applied to experimental spectra to calculate the specific activity (Bq/gm) of the sample of irregular shape

Part of:
Proceedings of the thirty-third IARP international conference on developments towards improvement of radiological surveillance at nuclear facilities and environment: book of abstracts

Additional details

Publishing Information

Publisher
Indian Association for Radiation Protection
Imprint Place
Mumbai (India)
Imprint Title
Proceedings of the thirty-third IARP international conference on developments towards improvement of radiological surveillance at nuclear facilities and environment: book of abstracts
Imprint Pagination
314 p.
Journal Page Range
p. 241

Conference

Title
33. IARP international conference on developments towards improvement of radiological surveillance at nuclear facilities and environment
Acronym
IARPIC-2018
Dates
16-20 Jan 2018
Place
Mumbai (India)

Optional Information

Notes
1 ref., 2 figs., 1 tab.