Published April 15, 1985 | Version v1
Journal article

Novel measuring strategies in neutron interferometry

  • 1. Dortmund Univ. (Germany, F.R.). Inst. fuer Physik
  • 2. Institut Max von Laue - Paul Langevin, 38 - Grenoble (France). Centre de Tri

Description

Angular misalignment of a sample in a single crystal neutron interferometer leads to systematic errors of the effective sample thickness and in this way to errors in the determination of the coherent scattering length. The misalignment can be determined and the errors can be corrected by a second measurement at a different angular sample position. Furthermore, a method has been developed which allows supervision of the wavelength during the measurements. These two techniques were tested by determining the scattering length of copper. A value of bsub(c) = 7.66(4) fm was obtained which is in excellent agreement with previous measurements. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods Phys. Res., Sect. A
Journal Volume
235
Journal Issue
3
Series
CODEN: NIMRD.;Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
557-560
ISSN
0168-9002