Published July 2003 | Version v1
Journal article

Intercomparison of dose distribution between Gamma Knife and C-arm-mounted linac

  • 1. Tokyo Univ. (Japan). Faculty of Medicine

Description

Although the Gamma Knife is widely used as a standard tool of radiosurgery for intracranial diseases, recent innovations in medical accelerators have actualized equivalent dose distribution. Most commonly, multiple non-coplanar arc beams through a combination of gantry and couch rotations are used in linac-based radiosurgery, but beam entry geometry is different from that with the Gamma Knife. We have developed a C-arm-mounted medical accelerator and realized conical beam entry without couch rotation. A combination of several conical rotation beams with different C-arm angles comprises multi-orbit dynamic conical radiation therapy (MODCRT) and resembles radiation beam entry with the Gamma Knife. Intercomparison of dose distributions and dose-volume histograms for the same brain tumor demonstrated that MODCRT delivers a similar dose fall-off pattern to that obtained with the Gamma Knife, though the low-dose spread patterns differ. (author)

Additional details

Publishing Information

Journal Title
Radiation Medicine
Journal Volume
21
Journal Issue
4
Journal Page Range
p. 178-182
ISSN
0288-2043